[Features] ●The microcomputer PID control method enables highly accurate temperature control with little setting deviation and overshoot. ●Equipped with an auto-start timer function as standard equipment. ●Auto start can be set for a maximum setting time of 99 hours and 59 minutes.
[Use] -
【specification】 - External dimensions (mm) width: 525 - External dimensions (mm) depth: 345 - External dimensions (mm) height: 500 ● Contents (L): 28 Power supply (V): Single phase 100 ●Temperature range (℃): 30 to 99.9 ●Safety device: earth leakage breaker, prevention of excessive temperature rise in tank, self-diagnosis function (heater disconnection, water level drop, temperature sensor abnormality, triac short) Inner tank dimensions (mm) width x depth x height: 350 x 300 x 300
[Specification 2] ●Temperature stability: ±0.02℃ Temperature control: Digital temperature indicator controller PID control method ●Temperature sensor: Platinum resistance type Pt100Ω - Stirring machine: vertical propeller stirring type 22W -Timer: Driving with auto start or auto stop timer (MAX 99 hours 59 minutes) ●Usable medium liquid: Water
[Material/Finish] -
[Set contents/accessories] -Stainless steel shelf plate and lid, 1 each
【Note】 -